Info

Richard Rainbolt

Richard Rainbolt
Microscopes Intl. LLC

April 20, 2018

Blog
Magnification and Resolution of Digital Microscopes

Power vs Pixels!

Nowhere is the magnification versus resolution question more prevalent than in digital microscopy. In many cases, these two terms are used interchangeably. However, they are distinct and should not be confused.

Magnification

Each objective lens has a magnification printed on its side. It's easy to understand. A 40x objective makes things appear 40 times larger than they actually are. Comparing objective magnification is relative—a 40x objective makes things twice as big as a 20x objective while a 60x objective makes them six times larger than a 10x objective.

The eyepiece in a typical desktop microscope is 10x. The product of the objective magnification and the eyepiece magnification gives the final magnification of the microscope. So, a 60x objective and a 10x eyepiece gives a total magnification of 600x.

So, what happens when you couple an objective to a 2 or 5 or 8-megapixel camera with no eyepiece? Then, what is the magnification? If you use a 20x objective, is the final image 20 times larger? 200 times larger?

Pixel Mapping

In digital microscopy, we use a term called pixel mapping to answer the question of digital image magnification. In general, a 20x objective maps 0.5 microns (of the specimen on the slide) to a single pixel on the camera. The final magnification is obtained by dividing the display pixel size (in microns) by the pixel mapping.

For a 70" HD TV (1920x1080), the pixel size is about 0.8mm (800 microns). And 800 divided by 0.5 gives a final magnification of 1,600x. For more magnification, you need a larger monitor!

Instead of displaying the image data pixel for pixel, we could display one image pixel on 4 (2x2), 9 (3x3) or 16 (4x4) display pixels and achieve double, triple, or quadruple the apparent magnification. But, that only gives us bigger pixels—not better resolution.

Resolution

Resolution is the objective's ability to resolve really small stuff. In objective terminology, this is the NA (numerical aperture) specification, which, like the magnification, is also printed on the side of the objective. The higher the NA, the more (smaller) stuff can be resolved. In general, higher magnification objectives have higher NA.

In general, for digital scanners, the maximum magnification of an objective is approximately 1000x the objective's NA. So, an objective with a .65NA can achieve approximately 650x in the digital domain. You might assume that we would always want the highest NA objective we could get, but that's not always the case.

Increased NA comes at a price: reduced depth of field and increased cost.

Numerical Aperture and Depth of Field are two sides of the same coin, and they are (more or less) inversely proportional. As NA increases, depth of field decreases, and vice-versa. Matching resolution and depth of field to the subject material is a key factor in choosing the right objective for the job at hand.

Depth of Field

Depth of Field is the distance between the nearest and farthest objects that are in focus without moving the objective. Anyone who regularly uses a desktop microscope knows that a 4x or 10x objective is much easier to focus than a 60x or 100x objective. The reason is because the depth of field of the lower NA objectives is very large (so, more stuff is in focus at the same time). And that means that getting the objective "close" to ideal focus is good enough because a deeper volume of the specimen is in good focus.

High NA objectives have a lower depth of field which means setting the focus is more critical and difficult to achieve. Getting the objective close-enough just doesn't work. Changing the focus only slightly can reveal different features in the specimen. Using an objective with a lower depth of field, objects may look different depending on how "deep" into the specimen you focus.

Striking a Balance

After all of the above detail, you might think it would be best to find a microscope objective with great magnification, high NA, and large depth of field. Such an objective does not exist—at any cost. So, we need to find a happy medium that satisfies as many requirements as possible. We make these kinds of trade-offs with whole slide scanners because of their general purpose nature.

How Does This Affect Whole Slide Scanning?

The best way to answer this question is to present a list of whole slide scanner truths to keep in mind.

  • Doubling the objective magnification quadruples the pixel data in the scan area. The reason for this is that the region doubles vertically and horizontally and we need to acquire four times as many images.
  • Doubling the magnification quadruples the scan time (more or less). Since you're capturing four times the image data, it will naturally take four times as long to do so.
  • Increasing the magnification (or NA) increases the precision (and time) required to focus. So, a scan with a 10x (low NA) objective is likely more than four (4) times faster than a 20x (high NA) objective. This is due to the greater precision required when focusing the higher NA objective. Focus gets even more critical with 40x or 60x objectives.

Conclusion

So, just what does all of this information and detail mean and what can we take away from it?

Choose the objective that matches the depth of field and resolution requirements of the job.

  • If you're looking at insects and most plant biology, a 10x or 20x objective is probably the right choice.
  • If you're looking at detail inside a white blood cell, you most likely require a 60x objective.
  • If you're looking at pathology, you may find that a 40x objective works well. But, a 20x objective may be just fine for many applications (the scans are certainly faster).

In any case, we are here to help. Just drop us an email or phone call to discuss your whole slide scanner requirements.